Online Multigas Gas analyzer
Off-Gas Analysis in Bioreactor Monitoring
AMB ELECTRONIC SYSTEM TECH PVT. LTD. Manufacturing Suitable multigas
analyzers to measure and monitor CO2, CH4, H2 and O2 simultaneously. The Off gas of the Bioreactor is injected to the gas analyzer by diffusion or by a sample gas pump provided inside the Panel. After proper flow, pressure regulation and filtration sample gas enter the Analyzer system at constant controlled Flow rate for further measurement.
Highly accurate multi technological detectors are Installed in the system used to measure different gas components in a single mixed gas and present them all at a time in a Single LCD screen. This is a Wall mounted compact system which includes the Gas analyzer, sample handling system as under, The analyzer Is required to be calibrate with standards and user can also verify at site when required and also able to calibrate for its Zero & Span just with a single touch.
As it is an analytical device , our this calibration feature is highly appreciated by user.
Then the system is connected to a PC running Data acquisition Soft wire to communicate and collect data from Gas analyzer and store the data for report generation in Excel sheet and trend generation. The Software is also enabled to display the real time data, trend on PC screen.
Online Multigas Gas analyzer with sample Handling system & Data Acquisition System
Consists of Wall mounting/ Floor standing dust-proof Control Panel enclose Containing Multi-Gas analyzer with Storage, Sample handling System and PC base Data Acquisition System running in windows10 for screen trend generation, report generation in excel and data storage.
Model No. : OGA-MA003
MAKE : AMB
Sample handling unit consists of Flow Meter with flow rate indication and control, Bypass flowmeter, safety valve, pressure regulators, filter, Gas cooling coil. Suitable Sample Pump provided to draw the sample from Source and inject to the analyzer via flow meter with control for continuous online measurement.
Detectors:
Gas analyzer Measuring and Monitoring CO2, CH4, H2, O2 simultaneously.
1. NDIR Detection Method Sensors used for CO2, CH4 Gases
2. Thermal conductivity Detection Method sensors used for H2 Gas
3. Electrochemical Detection Method and Sensor used for Oxygen Measurement
Range of measurement
1. CO2 Range :0.00- 100.00 %
2. CH4 Range :0.00- 100.00% %
3. H2 Range : 0.00- 100.00%
4. O2 Range : 0.00- 100.00%
5. Warm-up time :5 min (initial), <15 min
6. Sample flow meter with flow rate indication & Control : 0 -2 SCFH
7. Sample Port size : ¼ inches Swagelok type
8. Sample flow : by diffusion or by suitable Sample pump provide
9. Detection limit (3·σ) :< 0,5% F.S
10. Linearity :< ± 1% F.S.
11. Repeatability : ± 0.5% F.S.
12. Resolution : 0.1%
13. Operational Temp : 0-55 Deg C
14. Ambient humidity :0 – 95% rel. humidity (not condensing)
15. Response time(t90) :15s
16. Communication Port : USB / Ethernet for PC software connection
17. Power Supply : 98 to 240 VAC, 50 HZ
18. Display System : Color Touch TFT Screen for monitoring of all measured components
in One screen
& Control of calibration and sample pump using touch screen.
on screen Trend generation and report display in tabular format
19. Data acquisition & Storage :Inbuilt in analyzer without any PC
20. Calibration : one touch smart calibration for Zero & Span , with Last Calibration
Date & Time recorded on Screen.
21. Panel Size : 600 X600 X 500mm Approx
Table top Data Acquisition System Consists of following
22. Window 10 operating system
23. I3 processor, 8GB RAM
24. SCADA Software installed and running of PC to generate components wise Trends, Data Storage, report generation in excel.
Model No. : OGA-MA003 MAKE : AMB
Detectors:
Gas analyzer Measuring and Monitoring CO2, CH4, H2, O2
simultaneously
1. NDIR Detection Method Sensors for CO2, CH4 Gases
2. Thermal conductivity Detection Method sensors for H2 Gas
3. Electrochemical Detection Method Sensorfor Oxygen
Range of measurement
1. CO2 Range :0.00- 100.00 %
2. CH4 Range :0.00- 100.00% %
3. H2 Range : 0.00- 100.00%
4. O2 Range : 0.00- 100.00%
Specification.
1. Warm-up time :5 min (initial), <15 min
2. Sample flow meter with flow rate indication & Control : 0 -2 SCFH
3. Sample Port size : ¼ inches Swagelok type
4. Sample flow : by diffusion or by suitable Sample pump provide
5. Detection limit : < 0.5% F.S
6. Linearity :< ± 1% F.S.
7. Repeatability : ± 0.5% F.S.
8. Resolution : 0.1%
9. Operational Temp : 0-55 Deg C
10. Ambient humidity :0 – 95% rel. humidity (not condensing)
11. Response time(t90) :15s
12. Communication Port : USB / Ethernet for PC software connection
13. Power Supply : 98 to 240 VAC, 50 HZ
14. Display System : Color Touch TFT Screen for monitoring of all measured components
in one screen. Control of calibration & sample pump using touch screen. Onscreen Trend
generation and report display in tabular format.
Application: Fermentation gas. Bio gas, Bioreactor off gas